Coagulator Diatermo MB 122 - mono-bipolar 120W
                            High frequency electrosurgical unit suitable for monopolar or bipolar electrosurgery.
 Suitable for cutting (CUT), cut coagulation (BLEND), superficial coagulation (FORCED COAG), deep coagulation in absence of necrosis (SOFT COAG) and with forceps, bipolar coagulation.
 The use of distributed or normal neutral plate is possible.
 Applications:
  Leading edge technology, this advanced concept allows the device to be used for very professional operational procedures thanks to ergonomic and safety features.
 Open Circuit (OC) Alarm warns the user when the connection with the neutral plate isn't made, or if the cable is disconnected or the neutral plate is not connected with the patient. The alarm is shown by an intermittent beep together with a flashing red light on the front panel. In these cases, the high frequency for cutting and coagulation will be disabled.
 OVT Alarm, the unit is equipped with an internal circuit protection warning when the current operation is taking too long.
  
 Technical specifications:
  
 Minimal pre-selectable power: 0
 Level step: 1
 Maximum output power CUT (W): 120-250 ohm
 Maximum output power BLEND (W): 90-200 ohm
 Maximum output power FORCED COAG (W): 80-150 ohm
 Maximum output COAC SOFT (W): 60-100 ohm
 Maximum output power BIPOLAR (W): 40-100 ohm
 Crest Factor CUT: 1.5
 Crest Factor BLEND: 2.1
 Crest Factor COAG FORCED: 2.0
 Crest Factor COAG SOFT: 1.7
 Crest Factor BIPOLAR: 1.5
 Frequency: 600 kHz
 Weight: 7 kg
 LxHxD mm Size: 260 x 110 x 265
 Voltage (switchable): 115-230
 Frequency (Hz): 50-60
 Fuses 230Vac (5x20) Timed: 2x 3.15A
 Fuses 115Vac (5x20) Timed: 2x 6.3A
 Electrical input power (VA): 300
 Electrical input current (230Vac) (A): 1.3
 Electrical input current (115Vac) (A): 2.6
 Five steps adjustable sound level (from 55 to 75 dBA)
 Self-check monitoring.
 Power alert.
 Divided or normal plate.
 Remembers the last used settings.
 Electrical Class (EN60601 - 1) I CF
 MDD 93/42 / EEC Class: IIB
 Patient circuit: F
 Conforms to EN60601-1-4 (1998)
  
 Standard accessories ::
  
 Handle with two buttons single use.
 MB foot switch (to be used as an alternative to the grip)
 MB flat neutral plate and cable
 Kit of 6 short autoclavable electrodes, Ø 2,4 mm
 Power cable 2 m
 User Manual: English, Spanish, Italian, French, Polish
MDR